ACCESS ACCUTNI+3 REAGENT
Report
- Report Number
- 2122870-2015-00407
- Event Type
- Malfunction
- Date Received
- July 1, 2015
- Date of Event
- May 31, 2015
- Report Date
- June 9, 2015
- Manufacturer
- BECKMAN COULTER
- Product Code
- MMI
- PMA / PMN Number
- K121790
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS NOT DISPATCHED TO THE CUSTOMER'S SITE. THERE IS NO EVIDENCE THAT THE ACCESS ACCUTNI+3 REAGENT WAS RETURNED FOR EVALUATION. THE MARSEILLE COMPLAINT HANDLING UNIT (CHU) PERFORMED BOTH INTERFERENCE TESTING AND GEL FILTRATION CHROMATOGRAPHY (GFC) TESTING ON THE PATIENT'S SAMPLES. THE INTERFERENCE TESTING DID NOT DETECT EITHER A PATIENT SOURCE HETEROPHILIC OR ALKALINE PHOSPHATASE INTERFERENT IN THE PATIENT'S SAMPLES. GFC TESTING CONFIRMED THAT TROPONIN I WAS PRESENT IN THE PATIENT'S SAMPLES AND DID REACT ACCORDINGLY WITH THE ACCESS ACCUTNI+3 ASSAY. IN CONCLUSION, PRE-ANALYTICAL SAMPLE HANDLING MAY HAVE CONTRIBUTED TO THE ERRATIC ACCESS ACCUTNI+3 RESULTS OBTAINED BY THE CUSTOMER AS RE-CENTRIFUGATION OF THE SAMPLES GENERATED REPRODUCIBLE RESULTS. THE ACTUAL CAUSE OF THE EVENT CANNOT BE DETERMINED WITH THE AVAILABLE INFORMATION. (B)(6). ALL MDRS ASSOCIATED WITH THIS EVENT: MDR 2122870-2015-00404, MDR 2122870-2015-00405, MDR 2122870-2015-00406, MDR 2122870-2015-00407.
THE CUSTOMER REPORTED OBTAINING NON-REPRODUCIBLE TROPONIN I (ACCESS ACCUTNI+3) RESULTS FOR ONE (1) PATIENT INVOLVING THE LABORATORY'S UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM (SERIAL NUMBER (B)(4)). THE INITIAL ACCESS ACCUTNI+3 RESULT WAS ABOVE THE ASSAY'S NORMAL REFERENCE RANGE. THIS SAMPLE (SAMPLE 7) WAS THEN RE-CENTRIFUGED AND ANALYZED AN ADDITIONAL TWO (2) TIMES ON THE SAME UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM. THE CUSTOMER OBTAINED REPRODUCIBLE, LOWER RESULTS WHICH WERE STILL ABOVE THE ASSAY'S NORMAL REFERENCE RANGE. THE INITIAL, ELEVATED ACCUTNI+3 RESULTS WERE RELEASED FROM THE LABORATORY. THERE WAS NO REPORT OF PATIENT INJURY OR CHANGE IN PATIENT TREATMENT ASSOCIATED WITH THIS EVENT. MDR 2122870-2015-00404 WILL ADDRESS THE ERRATIC ACCESS ACCUTNI+3 RESULTS OBTAINED ON (B)(6) 2015 FOR THIS PATIENT ON THE LABORATORY'S ALTERNATE UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM (SERIAL NUMBER (B)(4)). MDR 2122870-2015-00405 WILL ADDRESS THE ERRATIC ACCESS ACCUTNI+3 RESULTS OBTAINED ON (B)(6) 2015 FOR THIS PATIENT ON THE LABORATORY'S ALTERNATE UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM AND MDR 2122870-2015-00406 WILL ADDRESS THE ERRATIC ACCESS ACCUTNI+3 RESULTS OBTAINED ON (B)(6) 2015 FOR THIS PATIENT. CALIBRATION, QC (QUALITY CONTROL) AND SYSTEM CHECK PARAMETERS WERE ALL RECOVERING WITHIN EXPECTED RANGES AT THE TIME OF THE EVENT. THE PATIENT'S SAMPLES WERE COLLECTED IN HEPARIN PLASMA TUBES. THE CUSTOMER DID NOT PROVIDE PROCESSING INFORMATION SUCH AS CENTRIFUGATION TIME AND SPEED. NO ISSUES WITH SAMPLE INTEGRITY, HARDWARE ERRORS OR SAMPLE FLAGS WERE REPORTED BY THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 425535 | ACCESS ACCUTNI+3 REAGENT | IMMUNOASSAY METHOD, TROPONIN SUBUNIT | MMI | BECKMAN COULTER | NA | 434885 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |