LIGACLIP CLIP APPLIER
Report
- Report Number
- 3005075853-2015-03953
- Event Type
- Malfunction
- Date Received
- July 1, 2015
- Date of Event
- June 22, 2015
- Report Date
- June 22, 2015
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION NOT AVAILABLE, DEVICE NOT RETURNED FOR ANALYSIS SHOULD THE INFORMATION BE PROVIDED LATER, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
(B)(4). ADDITIONAL INFORMATION: BATCH # M90969. CONCLUSION: THE ANALYSIS RESULTS FOUND THAT THE ER420 DEVICE WAS RETURNED WITH A CLIP IN THE JAWS; THE CLIP WAS REMOVED IN ORDER TO INSPECT THE JAWS AND THEY WERE FOUND TO BE IN GOOD CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. DURING THE ANALYSIS, THE DEVICE WAS CYCLED AND IT FED, RETAINED AND FORMED THE REMAINING CLIPS AS INTENDED. AS THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL, IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
(B)(4). BATCH # M90969. CONCLUSION: THE ANALYSIS RESULTS FOUND THAT THE ER420 DEVICE WAS RETURNED WITH A CLIP IN THE JAWS; THE CLIP WAS REMOVED IN ORDER TO INSPECT THE JAWS AND THEY WERE FOUND TO BE IN GOOD CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. DURING THE ANALYSIS, THE DEVICE WAS CYCLED AND IT FED, RETAINED AND FORMED THE REMAINING CLIPS AS INTENDED. AS THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL, IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THE CLIP WAS TWISTED. THE DEVICE WAS USED ON THE BLOOD VESSEL. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 426644 | LIGACLIP CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |