FDA Adverse Event Malfunction Summary report: N

LIGACLIP CLIP APPLIER

MDR report key: 4883309 · Received July 1, 2015

Report

Report Number
3005075853-2015-03953
Event Type
Malfunction
Date Received
July 1, 2015
Date of Event
June 22, 2015
Report Date
June 22, 2015
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION NOT AVAILABLE, DEVICE NOT RETURNED FOR ANALYSIS SHOULD THE INFORMATION BE PROVIDED LATER, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: BATCH # M90969. CONCLUSION: THE ANALYSIS RESULTS FOUND THAT THE ER420 DEVICE WAS RETURNED WITH A CLIP IN THE JAWS; THE CLIP WAS REMOVED IN ORDER TO INSPECT THE JAWS AND THEY WERE FOUND TO BE IN GOOD CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. DURING THE ANALYSIS, THE DEVICE WAS CYCLED AND IT FED, RETAINED AND FORMED THE REMAINING CLIPS AS INTENDED. AS THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL, IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). BATCH # M90969. CONCLUSION: THE ANALYSIS RESULTS FOUND THAT THE ER420 DEVICE WAS RETURNED WITH A CLIP IN THE JAWS; THE CLIP WAS REMOVED IN ORDER TO INSPECT THE JAWS AND THEY WERE FOUND TO BE IN GOOD CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. DURING THE ANALYSIS, THE DEVICE WAS CYCLED AND IT FED, RETAINED AND FORMED THE REMAINING CLIPS AS INTENDED. AS THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL, IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THE CLIP WAS TWISTED. THE DEVICE WAS USED ON THE BLOOD VESSEL. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
426644 LIGACLIP CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1