FDA Adverse Event Injury Summary report: N

PURELY YOURS

MDR report key: 4882688 · Received June 25, 2015

Report

Report Number
3009974348-2015-00097
Event Type
Injury
Date Received
June 25, 2015
Date of Event
May 9, 2015
Report Date
May 27, 2015
Manufacturer
AMEDA INC.
Product Code
HGX
PMA / PMN Number
K973501
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

PRODUCT WAS EVALUATED FOR EVIDENCE OF ALLEGATION. THE RETURNED AMEDA PURELY YOURS BREAST PUMP MET AMEDA SPECS FOR BOTH SUCTION AND SPEED, AND PASSED VISUAL INSPECTION STANDARDS. NO EVIDENCE OF MALFUNCTION WAS OBSERVED.

Description of Event or Problem · 1

CUSTOMER CONTACTED AMEDA, INC. ON (B)(6) 2015 TO REPORT LOW SUCTION AND DECREASED MILK OUTPUT WHILE USING THE PURELY YOURS BREAST PUMP THAT HAS BEEN ONGOING FOR WEEKS. CUSTOMER STATES SHE USED THE PURELY YOURS PUMP MORE FREQUENTLY PRIOR TO A DIAGNOSIS OF UNILATERAL RIGHT BREAST MASTITIS ON (B)(6) 2015. DUE TO BABY'S CONGESTION FROM A COLD AND HIS INABILITY TO NURSE, MOM USED THE PURELY YOURS PUMP EXCLUSIVELY EARLIER IN THE MONTH. CUSTOMER CONTACTED HER HEALTH CARE PROVIDER ON (B)(6)2015 AND WAS PRESCRIBED A 10 DAY COURSE OF ANTIBIOTICS. CUSTOMER FELT WELL WITHIN 3 DAYS SO SHE STOPPED THE ANTIBIOTICS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
410788 PURELY YOURS ELECTRIC BREAST PUMP HGX AMEDA INC. 24502082

Patients

Seq Age Sex Outcome Treatment
1 Other