FDA Adverse Event
Summary report: N
NCM SYSTEM
MDR report key: 488140
·
Received September 30, 2003
Report
- Report Number
- 2027454-2003-00006
- Date Received
- September 30, 2003
- Date of Event
- September 11, 2003
- Report Date
- September 11, 2003
- Manufacturer
- HILL-ROM MANUFACTURING, INC.
- Product Code
- HGM
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
COMLINX NCM SOFTWARE VERISON 3.1.0. ALLOWS AUDIBLE ANNUNCIATION, INCLUDING CODE BLUE ANNUNICATION, AT MINI-MASTER STATIONS TO BE DISABLED UNINTENTIONALLY. THIS ISSUE HAS NOT BEEN REPORTED WITH OTHER SOFTWARE RELEASES. THE INFO WAS COMMUNICATED TO HILL-ROM TECHNICAL SUPPORT FROM A HILL-ROM TECHNICIAN DURING THE INSTALLATION AND CERTIFICATION OF THE PRODUCT. NO ADVERSE OUTCOME TO PTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NCM SYSTEM | COMMUNICATIONS | HGM | HILL-ROM MANUFACTURING, INC. | P250001 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |