FDA Adverse Event Summary report: N

NCM SYSTEM

MDR report key: 488140 · Received September 30, 2003

Report

Report Number
2027454-2003-00006
Date Received
September 30, 2003
Date of Event
September 11, 2003
Report Date
September 11, 2003
Manufacturer
HILL-ROM MANUFACTURING, INC.
Product Code
HGM
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

COMLINX NCM SOFTWARE VERISON 3.1.0. ALLOWS AUDIBLE ANNUNCIATION, INCLUDING CODE BLUE ANNUNICATION, AT MINI-MASTER STATIONS TO BE DISABLED UNINTENTIONALLY. THIS ISSUE HAS NOT BEEN REPORTED WITH OTHER SOFTWARE RELEASES. THE INFO WAS COMMUNICATED TO HILL-ROM TECHNICAL SUPPORT FROM A HILL-ROM TECHNICIAN DURING THE INSTALLATION AND CERTIFICATION OF THE PRODUCT. NO ADVERSE OUTCOME TO PTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NCM SYSTEM COMMUNICATIONS HGM HILL-ROM MANUFACTURING, INC. P250001 NA

Patients

Seq Age Sex Outcome Treatment
1 NA