FDA Adverse Event Malfunction Summary report: N

SAVINA 300

MDR report key: 4875126 · Received June 9, 2015

Report

Report Number
9611500-2015-00091
Event Type
Malfunction
Date Received
June 9, 2015
Date of Event
April 2, 2015
Report Date
June 9, 2015
Manufacturer
DRAGER MEDICAL GMBH
Product Code
CBK
PMA / PMN Number
K121886
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FI
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION WAS STARTED BUT HAS NOT YET BEEN CONCLUDED. THE INVESTIGATION RESULT WILL BE REPORTED IN THE FOLLOW-UP REPORT.

Additional Manufacturer Narrative · 1

THE AVAILABLE INFORMATION HAS BEEN ANALYZED. TWO ERROR CODES IN THE DEVICE LOG (70 0005 AND 93 0005) WERE FOUND, WHICH POINT TO A TEMPORARY COMMUNICATION PROBLEM BETWEEN FRONT PROCESSOR AND MAINBOARD. THE DEVICE INTERNAL MONITORING DETECTED THE DATA INCONTINENCE AND TRIGGERED A RESTART TO RESTORE THE DATA BASE FOR VENTILATION. THE RESTART IS THE SPECIFIED BEHAVIOR FOR THE OBSERVED CONDITIONS. AS THE BEHAVIOR WAS NOT REPEATED, IT CAN BE CONCLUDED THAT NO HARDWARE FAILURE HAS CAUSED THE EVENT. ALSO NO SOFTWARE BUG IS KNOWN CONCERNING THIS ISSUE. DURING THE RESTART, WHICH NEEDS LESS THAN 12 SECONDS, THE PATIENT SYSTEM IS OPENED TO ATMOSPHERE TO ALLOW SPONTANEOUS BREATHING. THEREAFTER THE VENTILATION IS CONTINUED WITH THE PREVIOUS SETTINGS. THE DEVICE POSTS ALARM UNTIL THE LOWER ALARM LIMIT FOR MINUTE VOLUME HAS BEEN EXCEEDED AGAIN. IN THE FRAME OF OUR MARKET OBSERVATION THIS INTERNAL COMMUNICATION PROBLEM IS A RARE CASE AND NO FURTHER MEASURES ARE PLANNED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE VENTILATOR PERFORMED A RESTART DURING USE ON PT. AFTER 8 SECONDS VENTILATION WAS CONTINUED WITH THE PREVIOUS SETTINGS. NO INJURY WAS REPORTED.

Description of Event or Problem · 1

PLEASE SEE INITIAL REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
369888 SAVINA 300 VENTILATORS, INTENSIVE CARE CBK DRAGER MEDICAL GMBH UNK UNK

Patients

Seq Age Sex Outcome Treatment
1