FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 4874848 · Received June 29, 2015

Report

Report Number
2531779-2015-21419
Event Type
Malfunction
Date Received
June 29, 2015
Report Date
June 12, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
P130007
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED DAMAGED CASE. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2015. SUBMITTED: (B)(4) 2015. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2015 WITH THE FOLLOWING FINDINGS: DEVICE EVALUATION: ON INVESTIGATION, THE BATTERY COMPARTMENT WAS NOTED TO BE CRACKED AT THE CASE SEAL.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED DAMAGED CASE. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2015.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
420327 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1