AIR REAMER/DRILL
Report
- Report Number
- 8030965-2015-11076
- Event Type
- Malfunction
- Date Received
- June 27, 2015
- Report Date
- May 29, 2015
- Manufacturer
- SYNTHES OBERDORF
- Product Code
- HWE
- PMA / PMN Number
- PEXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(6). THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND OBSERVED THAT THE MOTOR WAS BLOCKED, SEIZED AND ROUGH RUNNING. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. IT WAS FURTHER DETERMINED THAT THE ROTOR AND BEARING WERE DEFECTIVE. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO WEAR FROM NORMAL USE AND SERVICING. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY.
IT WAS REPORTED FROM (B)(6) THAT THE MOTOR ON THE AIR REAMER/DRILL DEVICE DID NOT TURN. IT WAS NOT REPORTED IF THE DEVICE WAS USED IN SURGERY, OR IF THERE WAS PATIENT INVOLVEMENT. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS IN A SURGICAL PROCEDURE OR IF A SPARE DEVICE WAS AVAILABLE. IT WAS NOT REPORTED IF THERE WERE ANY INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. THE EXACT DATE OF THIS EVENT WAS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 417112 | AIR REAMER/DRILL | INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESSORIES AND ATTACHMENT | HWE | SYNTHES OBERDORF | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |