FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4871621 · Received June 26, 2015

Report

Report Number
3004209178-2015-69116
Event Type
Malfunction
Date Received
June 26, 2015
Date of Event
June 8, 2015
Report Date
June 8, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP RECEIVED WITH INTERMITTENT UP ARROW BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO UNEXPECTED NUMBERS RAMPING OR SCROLL BAR MOVING DURING TESTING. THE INSULIN PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, CRACKED CASE AT THE DISPLAY WINDOW CORNER, MINOR SCRATCHED LCD WINDOW, SCRATCHED RESERVOIR TUBE WINDOW, AND BROKEN BELT CLIP SLOT.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL INDICATING THAT THE INSULIN PUMP HAD KEYPAD ANOMALY. CUSTOMER'S BLOOD GLUCOSE WAS 198 MG.DL. THE CUSTOMER STATED THAT THE NUMBERS ARE RAMPING OR THE SCROLL BAR MOVING UP OR DOWN WITH NO INPUT FROM THE USER. THE CUSTOMER STATED THAT IT HAPPENED DURING NORMA USE. THE CUSTOMER WAS ADVISED THAT THE DEVICE WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
414718 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-754WWP

Patients

Seq Age Sex Outcome Treatment
1