FDA Adverse Event
Malfunction
Summary report: N
TRILOGY O2
MDR report key: 4871384
·
Received June 26, 2015
Report
- Report Number
- 2518422-2015-01815
- Event Type
- Malfunction
- Date Received
- June 26, 2015
- Date of Event
- June 2, 2015
- Report Date
- June 2, 2015
- Manufacturer
- RESPIRONICS, INC.
- Product Code
- CBK
- PMA / PMN Number
- K093905
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Description of Event or Problem · 1
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A "SERVICE REQUIRED" ALARM CONDITION OCCURRED AND THE DEVICE STOPPED OPERATING. THE DEVICE WAS NOT IN PATIENT USE. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, A "SERVICE REQUIRED" CODE WAS OBSERVED IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SYSTEM BOARD WAS REPLACED TO ADDRESS THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 414985 | TRILOGY O2 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS, INC. | 1040004 |