SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Report
- Report Number
- 1416980-2015-26794
- Event Type
- Malfunction
- Date Received
- June 26, 2015
- Date of Event
- June 16, 2015
- Report Date
- June 16, 2015
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AN INCOMPLETE PRIME WITH A CLOSED CLAMP ON THE PATIENT LINE DURING PRIMING IS THE KNOWN CAUSE OF THE AIR IN THE LINE. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO SYSTEMS PATIENT AT-HOME GUIDE¿, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. THE GUIDE INSTRUCTS THE USER TO OPEN THE CLAMP ON THE PATIENT LINE TO ENSURE PROPER PRIMING. IT PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET. IT WARNS THE USER NOT TO CONNECT TO THE PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A PATIENT EXPERIENCED AIR IN THE PATIENT LINE OF A HOMECHOICE CASSETTE WITHOUT AN ALARM. THIS OCCURRED DURING FILL ONE OF FIVE OF PERITONEAL DIALYSIS THERAPY. THE PATIENT WAS CONNECTED AT THE TIME OF THE EVENT. DURING TROUBLESHOOTING, IT WAS FOUND THAT THE PATIENT FORGOT TO OPEN THE CLAMP ON THE PATIENT LINE DURING PRIMING. THE TECHNICAL SERVICE REPRESENTATIVE ASSISTED THE PATIENT WITH ENDING THERAPY AND REVIEWED PROPER PROCEDURES AS PER USER MANUAL. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 415405 | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | HOMECHOICE |