FDA Adverse Event Malfunction Summary report: N

G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 4870138 · Received June 25, 2015

Report

Report Number
3004753838-2015-31483
Event Type
Malfunction
Date Received
June 25, 2015
Date of Event
May 30, 2015
Report Date
May 30, 2015
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. AN INTERIOR INSPECTION WAS PERFORMED AND THE INSPECTION PASSED. A REVIEW OF THE DOWNLOADED RECEIVER LOG DID NOT FIND ANY ISSUES RELATED TO THE CUSTOMER COMPLAINT. FUNCTIONAL TESTING AND A MANUAL DROP TEST WAS PERFORMED AND THE TESTS FAILED, CONFIRMING THE REPORTED EVENT OF NO AUDIO OUTPUT. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE SPEAKER.

Description of Event or Problem · 1

PATIENT'S MOTHER CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015 TO REPORT NO AUDIO OUTPUT FROM THE RECEIVER THAT OCCURRED ON (B)(6) 2015. AT THE TIME OF CONTACT, THE PATIENT'S MOTHER DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
412226 G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. MT22430-BLU 5197035

Patients

Seq Age Sex Outcome Treatment
1 13 YR