FDA Adverse Event
Malfunction
Summary report: N
G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM
MDR report key: 4870138
·
Received June 25, 2015
Report
- Report Number
- 3004753838-2015-31483
- Event Type
- Malfunction
- Date Received
- June 25, 2015
- Date of Event
- May 30, 2015
- Report Date
- May 30, 2015
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Additional Manufacturer Narrative · 1
(B)(4). THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. AN INTERIOR INSPECTION WAS PERFORMED AND THE INSPECTION PASSED. A REVIEW OF THE DOWNLOADED RECEIVER LOG DID NOT FIND ANY ISSUES RELATED TO THE CUSTOMER COMPLAINT. FUNCTIONAL TESTING AND A MANUAL DROP TEST WAS PERFORMED AND THE TESTS FAILED, CONFIRMING THE REPORTED EVENT OF NO AUDIO OUTPUT. THE ROOT CAUSE WAS DETERMINED TO BE A DEFECTIVE SPEAKER.
Description of Event or Problem · 1
PATIENT'S MOTHER CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015 TO REPORT NO AUDIO OUTPUT FROM THE RECEIVER THAT OCCURRED ON (B)(6) 2015. AT THE TIME OF CONTACT, THE PATIENT'S MOTHER DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 412226 | G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | MT22430-BLU | 5197035 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 13 YR |