FDA Adverse Event Malfunction Summary report: N

PIPELINE EMBOLIZATION DEVICE

MDR report key: 4869275 · Received June 25, 2015

Report

Report Number
2029214-2015-00730
Event Type
Malfunction
Date Received
June 25, 2015
Date of Event
June 3, 2015
Report Date
June 8, 2015
Manufacturer
COVIDIEN (IRVINE)
Product Code
OUT
PMA / PMN Number
P100018.S004
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

OFF LABEL USE: PIPELINE EMBOLIZATION DEVICE (PED) IS CONTRAINDICATED IN PATIENTS IN WHOM A PRE-EXISTING STENT IS IN PLACE IN THE PARENT ARTERY AT THE TARGET ANEURYSM LOCATION. IN THIS STUDY, THE PED WAS USED IN AN OFF-LABEL MANNER IN A PATIENT WITH PRE-EXISTING NITINOL SELF-EXPANDING STENTS. THE LOT HISTORY RECORD REVIEW WAS NOT POSSIBLE SINCE THE DEVICE LOT NUMBER WAS NOT REPORTED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS; THEREFORE, THE EVENT CAUSE COULD NOT BE DETERMINED. SERIOUS ADVERSE EVENTS FROM THE SAME ARTICLE WAS REPORTED IN MDR MFR# 2029214-2015-00731. (B)(4).

Description of Event or Problem · 1

MEDTRONIC (COVIDIEN) RECEIVED INFORMATION FROM LITERATURE REVIEW THAT ONE PIPELINE EMBOLIZATION DEVICE (PED) WAS NOT SUCCESSFULLY DEPLOYED. A LARGE FUSIFORM ANEURYSM OF THE BASILAR ARTERY COULD NOT BE TREATED, BECAUSE DURING STENT DEPLOYMENT, THE PED BECAME CAUGHT IN THE STRUTS OF THE EXISTING STENT, WHICH PREVENTED PROPER DEPLOYMENT AND LED TO AN ABORTION OF THE PROCEDURE. NO PATIENT INJURY WAS REPORTED AS A RESULT OF THIS PROCEDURE. IN THIS ARTICLE, THE AUTHORS REPORTED THEIR EXPERIENCE USING FLOW-DIVERTING STENTS IN THE REPAIR OF 25 ANEURYSMS FOR WHICH STENT-ASSISTED EMBOLIZATION HAD FAILED. CITATION: HEIFERMAN DM, BILLINGSLEY JT, KASLIWAL MK, ET AL. USE OF FLOW-DIVERTING STENTS AS SALVAGE TREATMENT FOLLOWING FAILED STENT-ASSISTED EMBOLIZATION OF INTRACRANIAL ANEURYSMS.J NEUROINTERV SURG. 2015 JUN 3. PII: NEURINTSURG-2015-011672.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
411429 PIPELINE EMBOLIZATION DEVICE INTRACRANIAL ANEURYSM FLOW DIVERTER OUT COVIDIEN (IRVINE)

Patients

Seq Age Sex Outcome Treatment
1 51 YR