BONE PIN- 3.2 X 140MM- STERILE (2 PACK)
Report
- Report Number
- 3005985723-2015-00111
- Event Type
- Injury
- Date Received
- June 25, 2015
- Date of Event
- June 18, 2015
- Report Date
- June 18, 2015
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- PMA / PMN Number
- K112507
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
REPORTED EVENT: DURING LEFT PARTIAL KNEE REPLACEMENT SURGERY, THE TIP OF A 3.2X140MM BONE PIN BROKE OFF AND EMBEDDED IN PATIENT AS SURGEON ATTEMPTED TO REMOVE THE ENTIRE PIN. DEVICE EVALUATION AND RESULTS: VISUAL INSPECTION SHOWED FAILURE IN TORSION THROUGH THE THREADS JUST ABOVE THE FLUTES. DEVICE HISTORY REVIEW: REVIEW OF THE DEVICE HISTORY RECORDS (SEE ATTACHMENT 1) FOR THE ASSOCIATED LOT INDICATED (B)(4) DEVICES (111621 NON-STERILE BONE PIN, SINGLE) WERE MANUFACTURED AND SHIPPED TO (B)(4) ON 05/15/2015 AND ACCEPTED INTO FINAL STOCK ON 05/15/2015 PER ERP. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN TRACKWISE AND CATSWEB RELATED TO P/N 111620, LOT NUMBER W39912 SHOWS ONE COMPLAINT RELATED TO THE FAILURE IN THIS INVESTIGATION. THIS COMPLAINT IS: (B)(4). TRACKING OF COMPLAINTS RELATED TO THE 111620 PART NUMBER WILL BE TRACKED THROUGH QUARTERLY TREND REQUEST #789. CONCLUSIONS: VISUAL INSPECTION CONFIRMED FAILURE OF THE BONE PIN. AS PART OF THE INVESTIGATION INTO THIS COMPLAINT, THE STRYKER REP THAT WAS PRESENT FOR THE CASE WAS CONTACTED. IT WAS DETERMINED THAT THE SURGEON DID NOT USE THE DRILL GUIDE DURING PLACEMENT OF BOTH BONE PINS. DUE TO THE NON-USE OF A DRILL GUIDE AS REQUIRED PER MAKOPLASTY PARTIAL KNEE APPLICATION USER GUIDE, 206388 REVISION 01 FOR PLACEMENT OF BONE PINS, THIS EVENT CONSTITUTES AN OFF-LABEL USE OF THE DEVICE. CORRECTIVE ACTION/PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD.
AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
THE SURGEON PERFORMED A PARTIAL KNEE ARTHROPLASTY PROCEDURE USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO) AND RESTORIS MULTICOMPARTMENTAL KNEE (MCK) IMPLANTS. DURING THE CASE, THE TIP OF THE BONE PIN BROKE OFF IN THE PATIENT'S TIBIA. THE SURGEON DECIDED TO LEAVE THE TIP OF THE BONE PIN IN THE PATIENT'S TIBIA. THE OUTCOME OF THE CASE WAS SUCCESSFUL.
THE SURGEON PERFORMED A PARTIAL KNEE ARTHROPLASTY PROCEDURE USING THE ROBOTIC ARM INTERACTIVE ORTHOPEDIC SYSTEM (RIO) AND RESTORIS MULTICOMPARTMENTAL KNEE (MCK) IMPLANTS. DURING THE CASE, THE TIP OF THE BONE PIN BROKE OFF IN THE PATIENT'S TIBIA. THE SURGEON DECIDED TO LEAVE THE TIP OF THE BONE PIN IN THE PATIENT'S TIBIA. THE OUTCOME OF THE CASE WAS SUCCESSFUL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 411739 | BONE PIN- 3.2 X 140MM- STERILE (2 PACK) | STEREOTAXIC DEVICE, ACCESSORY | OLO | MAKO SURGICAL CORP. | W39912 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |