STAT PROFILE M MGH SENSOR
MDR report key: 486714
·
Received September 30, 2003
Report
- Report Number
- 1219029-2003-00002
- Event Type
- Malfunction
- Date Received
- September 30, 2003
- Date of Event
- May 6, 2003
- Report Date
- September 29, 2003
- Manufacturer
- NOVA BIOMEDICAL
- Product Code
- CFA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
GENERAL MGR REPORTS THAT HOSP USING A STAT PROFILE M ANALYZER FOUND RECOVERY OF MG++ RESULTS OF PTS LOWER THAN A NOVA 8. THE AVERAGE LEVEL BY NOVA 8 WAS 0.52 MMOL/2 VS 0.35 MMOL/L ON STAT PROFILE M.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STAT PROFILE M MGH SENSOR | IONIZED MAGNESIUM SENSOR | CFA | NOVA BIOMEDICAL | I MAGNESIUM ELECTRODE | 207152 |