FDA Adverse Event Malfunction Summary report: N

G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

MDR report key: 4867114 · Received June 24, 2015

Report

Report Number
3004753838-2015-16184
Event Type
Malfunction
Date Received
June 24, 2015
Date of Event
May 29, 2015
Report Date
May 29, 2015
Manufacturer
DEXCOM, INC.
Product Code
MDS
PMA / PMN Number
P120005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SUBSEQUENT TO THE INITIAL MDR, THE RETURNED RECEIVER (PART NUMBER STK-DR-BLU /SERIAL NUMBER (B)(4) LOT NUMBER 5200549), WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION A REVIEW OF THE DOWNLOADED RECEIVER LOG DID NOT CONFIRM THE REPORTED EVENT OF INACCURACIES. A ROOT CAUSE COULD NOT BE DETERMINED.

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT DEVICE WASN'T RETURNED, BUT THE RECEIVER AND TRANSMITTER WAS USED WITH THE DEVICE HAS BEEN RECEIVED FOR EVALUATION ON (B)(4) 2015. A FOLLOW-UP REPORT WILL BE SUBMITTED ONCE THE EVALUATION IS COMPLETE. IT WAS REPORTED THAT THE PATIENT DID NOT CALIBRATE SINCE SEEING THE INACCURACY. IT SHOULD BE NOTED THAT THE DEXCOM G4® PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM USER'S GUIDE STATES: IF THE DIFFERENCE BETWEEN YOUR SENSOR GLUCOSE READING AND BLOOD GLUCOSE VALUE IS GREATER THAN 20% OF THE BLOOD GLUCOSE VALUE FOR SENSOR GLUCOSE READINGS > 80 MG/DL OR GREATER THAN 20 POINTS FOR SENSOR GLUCOSE READINGS < 80 MG/DL, WASH YOUR HANDS AND TAKE ANOTHER BLOOD GLUCOSE MEASUREMENT. IF THE DIFFERENCE BETWEEN THIS SECOND BLOOD GLUCOSE MEASUREMENT AND THE SENSOR IS STILL GREATER THAN 20% FOR SENSOR GLUCOSE READINGS > 80 MG/DL OR GREATER THAN 20 POINTS FOR SENSOR GLUCOSE READINGS < 80 MG/DL, RECALIBRATE YOUR SENSOR USING THE SECOND BLOOD GLUCOSE VALUE. THE SENSOR GLUCOSE READING WILL CORRECT OVER THE NEXT 15 MINUTES.

Description of Event or Problem · 1

PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015, TO REPORT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITORING (CGM) SYSTEM AND THE BLOOD GLUCOSE (BG) METER ON (B)(6) 2015. THE SENSOR WAS INSERTED ON (B)(6) 2015. PATIENT DID NOT CALIBRATE ACCORDING TO USER GUIDE RECOMMENDATIONS. THE PATIENT DID NOT REPORT INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
409444 G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM MDS MDS DEXCOM, INC. 9500-27 NI

Patients

Seq Age Sex Outcome Treatment
1 77 YR