FDA Adverse Event Injury Summary report: N

PURELY YOURS

MDR report key: 4864847 · Received June 23, 2015

Report

Report Number
3009974348-2015-00126
Event Type
Injury
Date Received
June 23, 2015
Date of Event
April 1, 2014
Report Date
May 14, 2014
Manufacturer
AMEDA, INC.
Product Code
HGX
PMA / PMN Number
K973501
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

PRODUCT WAS EVALUATED FOR EVIDENCE OF ALLEGATION. THE RETURNED AMEDA PURELY YOURS BREAST PUMP MET AMEDA SPECIFICATIONS FOR BOTH SUCTION AND SPEED, AND PASSED VISUAL INSPECTION STANDARDS. NO EVIDENCE OF MALFUNCTION WAS OBSERVED.

Description of Event or Problem · 1

AS PART OF AMEDA, INC.'S QUALITY MANAGEMENT SYSTEM ACTIVITIES, A REVIEW OF HISTORICAL COMPLAINTS WAS CONDUCTED. IT WAS DETERMINED THAT THIS COMPLAINT SHOULD HAVE BEEN REPORTED TO FDA. CUSTOMER CONTACTED AMEDA, INC ON (B)(6) 2014 TO REPORT THE PURELY YOURS BREAST PUMP SHE EXCLUSIVELY USES TO EXPRESS HER BREAST MILK HAS VERY LOW SUCTION AND IT NEVER FULLY DRAINS HER BREASTS OF MILK. CUSTOMER REPORTS DECREASE MILK OUTPUT AND SUBSEQUENT UNILATERAL RIGHT BREAST MASTITIS ON (B)(6) 2014. CUSTOMER CONTACTED HER HEALTH CARE PROVIDER AND WAS PRESCRIBED A 10 DAY COURSE OF ORAL ANTIBIOTICS TO RESOLVE MASTITIS. SHE BEGAN USING A MANUAL BREAST PUMP TO FULLY DRAIN HER BREASTS. MASTITIS RESOLVED WITHIN 1-2 DAYS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
406331 PURELY YOURS ELECTRIC BREAST PUMP, 884.5160 HGX AMEDA, INC. 24502082

Patients

Seq Age Sex Outcome Treatment
1 Other