FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4862672 · Received June 23, 2015

Report

Report Number
3004209178-2015-67839
Event Type
Malfunction
Date Received
June 23, 2015
Date of Event
June 4, 2015
Report Date
June 4, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

PUMP RECEIVED WITH NORMAL OPERATING CURRENTS. NO UNEXPECTED FAILED BATTERY TEST ALARM NOTED. ALL BUTTONS FUNCTIONING PROPERLY. HOWEVER, FOUND CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM DURING TESTING. ALSO RECEIVED WITH CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHED LCD WINDOW AND CRACKED BATTERY TUBE THREADS.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL INDICATING THAT THE INSULIN PUMP ALARM BUTTON ERROR. CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN MG/DL. THE CUSTOMER STATED THAT NONE OF THE BUTTONS WORK. THE CUSTOMER STATED THAT THE DEVICE WOULD BE REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO A BACKUP PLAN. THE CUSTOMER ALSO REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD A KEYPAD ANOMALY. CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN MG/DL. THE CUSTOMER STATED SHE WAS TRYING TO BOLUS AND THE NUMBERS WERE SCROLLING. CUSTOMER WAS ADVISED THAT THE DEVICE WOULD REPLACED AND AGREED TO RETURN THE PRODUCT FOR ANALYSIS. THE CUSTOMER DECLINED TO TROUBLESHOOT FOR BATTERY ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
407149 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAS

Patients

Seq Age Sex Outcome Treatment
1 49 YR