FDA Adverse Event
Malfunction
Summary report: N
VERSACARE BED
MDR report key: 4862563
·
Received June 18, 2015
Report
- Report Number
- 1824206-2015-00664
- Event Type
- Malfunction
- Date Received
- June 18, 2015
- Date of Event
- May 26, 2015
- Report Date
- May 26, 2015
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECHNICIAN FOUND THE EXTERNAL ALARM NEEDED TO BE REPLACED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED FROM 2012-2015. THE TECH REPLACED THE EXTERNAL ALARM TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem · 1
HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECH STATING THE BRAKES NOT SET ALARM DID NOT WORK. THE BED WAS LOCATED ON 3 TOWER AT THE ACCOUNT. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 398051 | VERSACARE BED | A/C POWERED ADJUSTABLE HOSP BED | FNL | HILL-ROM, INC. | 3200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |