FDA Adverse Event Malfunction Summary report: N

GOLVO 8008 LOWBASE WITH CABLE

MDR report key: 4862555 · Received June 18, 2015

Report

Report Number
8030916-2015-00081
Event Type
Malfunction
Date Received
June 18, 2015
Date of Event
May 26, 2015
Report Date
May 26, 2015
Manufacturer
LIKO AB
Product Code
FSA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACCOUNT CALLED HILL-ROM TECHNICAL SUPPORT TO TROUBLE SHOOT. THE ACCOUNT PLACED AN ORDER FOR A MIDDLE BEAM. THE GEAR RACK IS INCLUDED AS PART OF THE MIDDLE BEAM. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS LIFT. IT IS UNK IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR LIFTS. THE ACCOUNT WILL REPLACE THE MIDDLE BEAM WITH GEAR RACK SET TO RESOLVE THE ISSUE. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM REC'D A REPORT FROM THE ACCOUNT STATING THAT ONE LEG WILL NOT RETRACT. THE LIFT WAS LOCATED IN THE PT ROOM. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
397202 GOLVO 8008 LOWBASE WITH CABLE NON-AC POWERED PATIENT LIFT FSA LIKO AB P2000039

Patients

Seq Age Sex Outcome Treatment
1