FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4862361 · Received June 23, 2015

Report

Report Number
2032227-2015-19835
Event Type
Malfunction
Date Received
June 23, 2015
Date of Event
June 3, 2015
Report Date
June 3, 2015
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED FOR A BUTTON ERROR DUE TO MOISTURE DAMAGE TO THE KEYPAD TRACES. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW, CRACKED BATTERY TUBE THREADS, AND A CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

CUSTOMER REPORTED VIA PHONE, THE INSULIN PUMP WAS ALARMING BUTTON ERROR. CUSTOMER DIDN'T KNOW HIS BLOOD GLUCOSE AT THE TIME OF THE EVENT. CUSTOMER STATED THE DEVICE WAS ADMINISTERING A NORMAL BASAL WHEN IT ALARMED. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE, REVERT TO BACK UP PLAN, AND THE DEVICE NEED TO BE REPLACED. HE AGREED TO RETURN THE DEVICE FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
405715 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523LNAB

Patients

Seq Age Sex Outcome Treatment
1 9 YR