FDA Adverse Event
Malfunction
Summary report: N
E SERIES DEFIBRILLATOR
MDR report key: 4862091
·
Received June 16, 2015
Report
- Report Number
- 1220908-2015-01471
- Event Type
- Malfunction
- Date Received
- June 16, 2015
- Report Date
- May 27, 2015
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- MKJ
- PMA / PMN Number
- K042007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION AND THE CUSTOMER'S REPORT WAS NOT REPLICATED OR CONFIRMED. THE DEVICE WAS PUT THROUGH EXTENSIVE TESTING WITHOUT DUPLICATING THE REPORT. A SYSTEM INTERCONNECTION CABLE WAS REPLACED AS A PRECAUTION. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. NO TREND IS ASSOCIATED WITH REPORTS OF THIS TYPE.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT WHILE ATTEMPTING TO TREAT A PATIENT, THE DEVICE'S DISPLAY INTERMITTENTLY BLANKED OUT. COMPLAINANT INDICATED THAT THERE WAS NO ADVERSE EFFECT TO THE PATIENT DUE TO THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 392711 | E SERIES DEFIBRILLATOR | E SERIES | MKJ | ZOLL MEDICAL CORPORATION | E SERIES | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |