FDA Adverse Event
Malfunction
Summary report: N
CARE ASSIST BED
MDR report key: 4862049
·
Received June 16, 2015
Report
- Report Number
- 9615739-2015-00198
- Event Type
- Malfunction
- Date Received
- June 16, 2015
- Date of Event
- May 22, 2015
- Report Date
- May 22, 2015
- Manufacturer
- HILL-ROM PLUVIGNER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECHNICIAN FOUND BOTH BRAKE RODS WERE DAMAGED. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE BRAKE RODS TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem · 1
HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKES WERE NOT HOLDING. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).