SYSTEM 7 DUAL TRIGGER ROTARY
Report
- Report Number
- 0001811755-2015-02284
- Event Type
- Malfunction
- Date Received
- June 23, 2015
- Date of Event
- May 27, 2015
- Report Date
- May 27, 2015
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HWE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- OTHER
Narratives
THE REPORTED EVENT WAS CONFIRMED. SERVICE EVALUATION FOUND THAT THE BOTTOM TRIGGER WOULD STAY IN THE RUN POSITION CAUSING DEVICE RUN-ON, DUE TO CORROSION. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED. FAILURE ANALYSIS IS IN PROGRESS.
IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED BY A MANUFACTURER FIELD SERVICE TECHNICIAN AT THE USER FACILITY, THE DEVICE RAN WITHOUT USER ACTIVATION AFTER THE TRIGGER WAS RELEASED. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
IT WAS REPORTED THAT DURING EQUIPMENT TESTING CONDUCTED BY A MANUFACTURER FIELD SERVICE TECHNICIAN AT THE USER FACILITY, THE DEVICE RAN WITHOUT USER ACTIVATION AFTER THE TRIGGER WAS RELEASED. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.