G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2015-16112
- Event Type
- Malfunction
- Date Received
- June 22, 2015
- Date of Event
- May 20, 2015
- Report Date
- May 27, 2015
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE COMPLAINT DEVICE WAS NOT RETURNED FOR EVALUATION AND DATA WAS NOT PROVIDED; THEREFORE, THE REPORTED COMPLAINT OF INACCURACIES CANNOT BE CONFIRMED. IT WAS REPORTED THAT THE PATIENT DID NOT CALIBRATE ACCORDING TO THE USER'S GUIDE. IT SHOULD BE NOTED THAT THE DEXCOM G4® PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM USER'S GUIDE STATES: "DO NOT CALIBRATE IF YOUR BLOOD GLUCOSE IS CHANGING AT A SIGNIFICANT RATE, TYPICALLY MORE THAN 2 MG/DL PER MINUTE. DO NOT CALIBRATE WHEN YOUR RECEIVER SCREEN IS SHOWING THE RISING SINGLE ARROW OR DOUBLE ARROW, WHICH INDICATES THAT YOUR BLOOD GLUCOSE IS RISING 2-3 MG/DL/MIN OR MORE THAN 3 MG/DL/MIN. ALSO, DO NOT CALIBRATE WHEN YOUR RECEIVER SCREEN IS SHOWING THE FALLING SINGLE ARROW OR DOUBLE ARROW, WHICH INDICATES THAT YOUR BLOOD GLUCOSE IS FALLING 2-3 MG/DL/MIN OR MORE THAN 3 MG/DL/MIN. CALIBRATING DURING SIGNIFICANT RISE OR FALL OF BLOOD GLUCOSE MAY AFFECT ACCURACY OF SENSOR GLUCOSE READINGS. DO NOT CALIBRATE IF THE OUT OF RANGE SYMBOL SHOWS IN THE STATUS AREA. DO NOT CALIBRATE IF THE GLUCOSE READING ERROR SYMBOL SHOWS IN THE STATUS AREA."
PATIENT'S FATHER CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015, TO REPORT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITORING (CGM) SYSTEM AND THE BLOOD GLUCOSE (BG) METER ON (B)(6) 2015. THE SENSOR WAS INSERTED ON (B)(6) 2015. PATIENT DID NOT CALIBRATE ACCORDING TO USER GUIDE RECOMMENDATIONS. THE PATIENT'S FATHER DID NOT REPORT INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 403581 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | 9500-27 | 5198073 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 3 YR |