530G INSULIN PUMP
Report
- Report Number
- 3004209178-2015-67750
- Event Type
- Malfunction
- Date Received
- June 22, 2015
- Date of Event
- June 3, 2015
- Report Date
- June 3, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- OTHER
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE PUMP WAS RECEIVED WITH A CONSTANT MOTOR ERROR ALARM DURING THE REWIND DUE TO AN OUT OF PHASE MOTOR ENCODER SIGNAL. UNABLE TO PERFORM THE DISPLACEMENT TEST DUE TO THE CONSTANT MOTOR ERROR ALARM. THE PUMP WAS ALSO RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS AND MINOR SCRATCHES ON THE LCD WINDOW.
THE CUSTOMER REPORTED VIA PHONE CALL THAT SHE WAS GETTING A MOTOR ERROR ALARM ON THE INSULIN PUMP. CUSTOMER WAS GETTING A REGULAR BASAL DELIVERY WHEN THE ALARM OCCURRED. CUSTOMER'S BLOOD GLUCOSE WAS 247 MG/DL. CUSTOMER WAS ASSISTED WITH CLEARING THE ALARM. CUSTOMER USES THE SENSOR FEATURE ON THE PUMP. IT WAS EXPLAINED THAT A FALSE MOTOR ERROR MAY BE CAUSED BY VIEWING THE SENSOR GLUCOSE GRAPH WHILE THE PUMP IS DELIVERING A BOLUS. CUSTOMER WAS ADVISED THAT THEY CAN REDUCE THE LIKELIHOOD OF THESE OCCURRENCES BY ALLOWING THE BOLUS DELIVERY TO COMPLETE BEFORE ACCESSING THE SENSOR GLUCOSE GRAPH. CUSTOMER STATED THAT THEY ARE UNABLE TO COMPLETE THE REWIND SEQUENCE. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WILL NEED TO BE REPLACED. CUSTOMER WAS ALSO ADVISED TO REVERT TO A BACK-UP PLAN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 403960 | 530G INSULIN PUMP | OZO | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-551NAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR |