FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4860674 · Received June 22, 2015

Report

Report Number
3004209178-2015-67716
Event Type
Malfunction
Date Received
June 22, 2015
Date of Event
June 2, 2015
Report Date
June 3, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RECEIVED WITH NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM NOTED DURING TESTING. THE PUMP ALSO HAD A CRACKED RESERVOIR TUBE LIP, A CRACKED CASE NEAR THE DISPLAY WINDOW CORNER, A CRACKED DISPLAY WINDOW, AND CRACKED BATTERY TUBE THREADS.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

A CUSTOMER'S MOTHER REPORTED VIA PHONE CALL INDICATING THE CUSTOMER'S INSULIN PUMP ALARMED BUTTON ERROR. THE PATIENT'S BLOOD GLUCOSE LEVEL WAS 300 MG/DL AT THE TIME OF THE CALL. THE CUSTOMER STATED THAT THERE WERE NO SIGNIFICANT EVENTS THAT COULD HAVE LED TO THE ISSUE. THE CUSTOMER WAS INFORMED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK UP PLAN. THE CUSTOMER SENT THE PRODUCT BACK FOR ANALYSIS. A REPLACEMENT PUMP WAS SHIPPED TO THE CUSTOMER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
404092 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722WWS

Patients

Seq Age Sex Outcome Treatment
1