FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 4860576
·
Received June 22, 2015
Report
- Report Number
- 3004209178-2015-67630
- Event Type
- Malfunction
- Date Received
- June 22, 2015
- Date of Event
- June 3, 2015
- Report Date
- June 3, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Additional Manufacturer Narrative · 1
THE PUMP WAS RECEIVED WITH A CRACKED AND BLEEDING GLASS. UNABLE TO TEST FOR A BLANK DISPLAY DUE THE CRACKED AND BLEEDING GLASS. THE PUMP ALSO HAD MINOR SCRATCHES ON THE LCD WINDOW.
Description of Event or Problem · 1
CUSTOMER REPORTED VIA PHONE CALL TO HAVE A BLANK DISPLAY SCREEN. CUSTOMER'S BLOOD GLUCOSE WAS UNKNOWN. AFTER TROUBLESHOOTING, DISPLAY SCREEN DID NOT RETURN. CUSTOMER WAS ADVISED THAT INSULIN PUMP WOULD NEED TO BE REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 404270 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723LNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR |