FDA Adverse Event Malfunction Summary report: N

RESERVOIR 3ML

MDR report key: 4860549 · Received June 22, 2015

Report

Report Number
3004209178-2015-67595
Event Type
Malfunction
Date Received
June 22, 2015
Date of Event
June 3, 2015
Report Date
June 3, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K032005
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSPECTED RESERVOIR, SNAP-CAP AND SEPTUM FOR ANOMALIES NONE WERE FOUND. PERFORMED OCCLUSION TEST PER SPECIFICATIONS, RESERVOIR FILLED AND CONNECTED TO NEW INFUSION SET. INSTALLED RESERVOIR AND CONNECTOR INTO INSULIN PUMP AND PERFORMED INFUSION SET. INSTALLED RESERVOIR AND CONNECTOR TO INSULIN PUMP AND PERFORMED CATHETER TIP. RESERVOIR NOT OCCLUDED.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

A CUSTOMER REPORTED VIA PHONE CALL INDICATING THAT THEIR INSULIN PUMP ALARMED NO DELIVERY. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 147 MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER STATED THAT THERE WERE NO SIGNIFICANT EVENTS THAT COULD HAVE LED TO THE ISSUE. THE CUSTOMER WAS ASSISTED WITH TROUBLE SHOOTING AND FOUND THAT THE RESERVOIR NEEDED TO BE CHANGED. THE ISSUE WAS RESOLVED WITH A SET CHANGE. THE CUSTOMER SENT THE SETS BACK FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
403153 RESERVOIR 3ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-332A

Patients

Seq Age Sex Outcome Treatment
1 74 YR