FDA Adverse Event Injury Summary report: N

GORE® TAG® THORACIC ENDOPROSTHESIS

MDR report key: 4858287 · Received June 21, 2015

Report

Report Number
2017233-2015-00374
Event Type
Injury
Date Received
June 21, 2015
Date of Event
June 29, 2011
Report Date
May 26, 2015
Manufacturer
W.L. GORE & ASSOCIATES
Product Code
MIH
PMA / PMN Number
P040043
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE REVIEW OF THE MANUFACTURING PAPERWORK VERIFIED THAT THIS LOT MET ALL PRE-RELEASE SPECIFICATIONS. USERS ARE MADE AWARE OF THE RISKS ASSOCIATED WITH TYPE II ENDOLEAKS IN THE IFU AND ARE INSTRUCTED TO CONSIDER THE RISKS AND BENEFITS DISCUSSED IN THE IFU FOR EACH PATIENT BEFORE USING THE DEVICES.

Description of Event or Problem · 1

ON (B)(6) 2010, THE PATIENT UNDERWENT AORTIC ARCH REPLACEMENT USING A GRAFT. ON (B)(6) 2010, THE PATIENT UNDERWENT ENDOVASCULAR REPAIR OF A THORACIC AORTIC ANEURYSM USING TWO GORE® TAG® THORACIC ENDOPROSTHESES (TGT3720/8317587 AND TGT4015/8233020). TWO STENT GRAFTS WERE DEPLOYED FROM THE DISTAL END OF THE EXISTING GRAFT USING AN ELEPHANT-TRUNK APPROACH. THE PROCEDURE WAS CONCLUDED WITHOUT ENDOLEAKS PRESENT. ON (B)(6) 2010, THE PATIENT WAS DISCHARGED OF THE HOSPITAL. ANEURYSM DIAMETER WAS 68MM, AND NO ENDOLEAKS HAD BEEN PRESENT. ON (B)(6) 2011, IN SIX-MONTH FOLLOW-UP STUDY, A TYPE II ENDOLEAK HAD BEEN REVEALED. ANEURYSM DIAMETER WAS 68MM. ON (B)(6) 2011, IN ONE-YEAR FOLLOW-UP STUDY, THE TYPE II HAD PERSISTED. ANEURYSM DIAMETER HAD REMAINED UNCHANGED. A WAIT-AND-WATCH APPROACH WAS TAKEN TO THE ENDOLEAK. ON (B)(6) 2012, IN TWO-YEAR FOLLOW-UP STUDY, ANEURYSM DIAMETER HAD ENLARGED TO 70MM WITH PERSISTENT TYPE II ENDOLEAK. REINTERVENTION WAS NOT PERFORMED. ON (B)(6) 2013, IN THREE-YEAR FOLLOW-UP STUDY, ANEURYSM DIAMETER HAD STILL ENLARGED TO 77MM WITH THE PERSISTENT TYPE II ENDOLEAK. A WAIT-AND-WATCH APPROACH HAD BEEN CONTINUED TO THE ENDOLEAK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
401998 GORE® TAG® THORACIC ENDOPROSTHESIS SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH W.L. GORE & ASSOCIATES 8317587

Patients

Seq Age Sex Outcome Treatment
1 79 YR Hospitalization TGT4015/8233020