PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2015-66596
- Event Type
- Malfunction
- Date Received
- June 18, 2015
- Date of Event
- May 30, 2015
- Report Date
- May 30, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT
- Reporter Occupation
- PATIENT
Narratives
THE PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM WAS NOTED. NO SCROLLING NUMBERS ANOMALY NOTED. THE PUMP ALSO HAD CRACKED BATTERY TUBE THREADS, A CRACKED RESERVOIR TUBE LIP, AND A MISSING END CAP STICKER.
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
A CUSTOMER REPORTED VIA PHONE CALL INDICATING THAT THEIR INSULIN PUMP ALARMED BUTTON ERROR. THE PATIENT'S BLOOD GLUCOSE LEVEL WAS 120 MG/DL AT THE TIME OF THE CALL. THE CUSTOMER STATED THAT THERE WERE NO SIGNIFICANT EVENTS THAT COULD HAVE LED TO THE ISSUE. THE CUSTOMER WAS INFORMED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK UP PLAN. THE CUSTOMER SENT THE PRODUCT BACK FOR ANALYSIS. A REPLACEMENT PUMP WAS SHIPPED TO THE CUSTOMER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 398203 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR |