FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4853486 · Received June 18, 2015

Report

Report Number
2531779-2015-20257
Event Type
Malfunction
Date Received
June 18, 2015
Report Date
June 3, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/22/2015 WITH THE FOLLOWING FINDINGS: THERE WAS EVIDENCE OF CS 064 OBSERVED IN THE BLACK BOX DATA. THE PUMP WAS EXERCISED FOR 24 HOURS AND THE COMPLAINT WAS NOT DUPLICATED. THE PUMP CASE WAS REMOVED AND THERE WERE NO DEFECTS FOUND INTERNALLY.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CS 064) ISSUE. IT WAS REPORTED THAT A CS-064 ¿CALL SERVICE ALARM¿ OCCURRED DURING THE REWIND/LOAD STEP. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
398549 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 70 YR