ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2015-20257
- Event Type
- Malfunction
- Date Received
- June 18, 2015
- Report Date
- June 3, 2015
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/22/2015 WITH THE FOLLOWING FINDINGS: THERE WAS EVIDENCE OF CS 064 OBSERVED IN THE BLACK BOX DATA. THE PUMP WAS EXERCISED FOR 24 HOURS AND THE COMPLAINT WAS NOT DUPLICATED. THE PUMP CASE WAS REMOVED AND THERE WERE NO DEFECTS FOUND INTERNALLY.
ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CS 064) ISSUE. IT WAS REPORTED THAT A CS-064 ¿CALL SERVICE ALARM¿ OCCURRED DURING THE REWIND/LOAD STEP. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 398549 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR |