FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4853283 · Received June 18, 2015

Report

Report Number
3004209178-2015-66514
Event Type
Malfunction
Date Received
June 18, 2015
Date of Event
May 30, 2015
Report Date
May 30, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTONS DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED. NO FROZEN SCREEN NOTED. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON LCD WINDOW, CRACKED BATTERY TUBE THREADS AND BROKEN RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED VIA PHONE CALL THAT SHE HAD A BUTTON ERROR ALARM ON THE INSULIN PUMP. CUSTOMER'S BLOOD GLUCOSE AT THE TIME WAS 137 MG/DL. CUSTOMER STATED THAT THE PUMP JUST STOPPED WORKING AND THE PUMP WAS SITTING ON HER WAIST WHEN SHE NOTICED THE BUTTON ERROR. CUSTOMER STATED THAT THE SCREEN WAS FROZEN FOR 5 MINUTES AND THEN THE BUTTON ERROR OCCURRED. CUSTOMER DECLINED TROUBLESHOOTING FOR THE FROZEN DISPLAY. CUSTOMER WAS ADVISED THAT THE PUMP WILL NEED TO BE REPLACED. CUSTOMER WAS ALSO ADVISED TO DISCONTINUE USE OF THE PUMP AND REVERT TO A BACK-UP PLAN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
397309 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAS

Patients

Seq Age Sex Outcome Treatment
1 35 YR