FDA Adverse Event Malfunction Summary report: N

LIGACLIP CLIP APPLIER

MDR report key: 4852816 · Received June 18, 2015

Report

Report Number
3005075853-2015-03704
Event Type
Malfunction
Date Received
June 18, 2015
Date of Event
June 1, 2015
Report Date
June 1, 2015
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: THE ER320 DEVICE WAS RETURNED FOR ANALYSIS AND UPON INSPECTION THE JAWS WERE FOUND TO BE IN A YIELDED CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT THE DEVICE WAS FUNCTIONALLY EVALUATED. UPON FIRING OF THE DEVICE, THE REMAINING CLIPS WERE DROPPED DUE TO THE CONDITION OF THE JAWS; FINALLY THE INSTRUMENT LOCKED OUT AS INTENDED. POSSIBLE CAUSES FOR THE FOUND CONDITION OF THE YIELDED JAWS MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE FIRST 3 CLIPS FIRED AS EXPECTED. SUBSEQUENT CLIPS ALL CAME OUT SCISSORED AND HAD TO BE REMOVED OFF VESSEL. ANOTHER ER320 WAS OPENED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
396794 LIGACLIP CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA M4HD43

Patients

Seq Age Sex Outcome Treatment
1