ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2015-20176
- Event Type
- Malfunction
- Date Received
- June 17, 2015
- Report Date
- June 4, 2015
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. USE ERROR SHOULD BE CONSIDERED AS TIME WAS INCORRECTLY SET.
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/20/2015 WITH THE FOLLOWING FINDINGS: REVIEW OF THE TDD HISTORY SHOWS THE USER SET THE TIME/DATE INCORRECTLY; 2 RECORDS FOR (B)(6) 2015 WERE RECORDED IN THE TDD HISTORY THE DAILY INSULIN DELIVERY TOTALS CORRECTLY REFLECT USER'S PROGRAMMED BASAL RATES. PUMP PASSED DELIVERY ACCURACY TEST AND FOUND TO BE DELIVERING WITHIN REQUIRED RANGE AND DELIVERING ACCURATELY. PUMP WAS EXERCISED FOR 24HRS; NO ALARMS WERE DUPLICATED. THE CONCLUSION OF THE INVESTIGATION WAS USER ERROR. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS AND ALLEGED THAT THE PATIENT HAD A BLOOD GLUCOSE (BG) OF 43 MG/DL WITH SYMPTOMS OF COGNITIVE IMPAIRMENT AND CONFUSION. DURING TROUBLESHOOTING, CUSTOMER SUPPORT FOUND THE PATIENT HAD SET THE TIME IN THE PUMP INCORRECTLY, TIME WAS SET FOR P.M. AS OPPOSED TO A.M. THE PATIENT REQUIRED NO UNUSUAL TREATMENT AND CONTINUES ON PUMP THERAPY. THIS COMPLAINT IS BEING REPORTED AS THE PATIENT EXPERIENCED HYPOGLYCEMIA FOLLOWING THE USE ERROR OF SETTING THE TIME IN THE PUMP INCORRECTLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 395554 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Life Threatening |