G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Report
- Report Number
- 3004753838-2015-15970
- Event Type
- Malfunction
- Date Received
- June 17, 2015
- Date of Event
- May 23, 2015
- Report Date
- May 23, 2015
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(6). SUBSEQUENT TO THE INITIAL MDR, THE COMPLAINT SENSOR WAS NOT RETURNED TO DEXCOM. THE RECEIVER ((B)(4) LOT NUMBER 5191734), BEING USED WITH THE COMPLAINT SENSOR, WAS RETURNED ON (B)(6) 2015. THE RETURNED RECEIVER WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONALTESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION; HOWEVER, A REVIEW OF THE DIAGNOSTIC CHART CONFIRMED THE REPORTED EVENT IF INACCURACIES. A DEFINITIVE ROOT CAUSE COULD NOT BE DETERMINED.
(B)(4). THE COMPLAINT DEVICE WAS NOT RETURNED FOR EVALUATION; HOWEVER, THE DATA LOG WAS PROVIDED BY THE PATIENT. IT WAS DOWNLOADED AND REVIEWED ON (B)(4) 2015 CONFIRMING THE REPORTED EVENT OF INACCURATE READINGS.
PATIENT CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2015, TO REPORT AN INACCURACY BETWEEN THE CONTINUOUS GLUCOSE MONITORING (CGM) SYSTEM AND THE BLOOD GLUCOSE (BG) METER ON (B)(6) 2015. THE SENSOR WAS INSERTED ON (B)(6) 2015. THE PATIENT DID NOT REPORT INJURY OR MEDICAL INTERVENTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 393892 | G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | 9500-27 | 5196043 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR |