FDA Adverse Event Malfunction Summary report: N

TRILOGY O2

MDR report key: 4851162 · Received June 17, 2015

Report

Report Number
2518422-2015-01714
Event Type
Malfunction
Date Received
June 17, 2015
Date of Event
May 25, 2015
Report Date
May 25, 2015
Manufacturer
RESPIRONICS, INC.
Product Code
CBK
PMA / PMN Number
K093905
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Description of Event or Problem · 1

THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR BEGAN OPERATING WITHOUT KEYPRESS ACTIVATION WHEN THE DEVICE WAS CONNECTED TO AC POWER. THE DEVICE WAS NOT IN PATIENT USE. THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED. THE DEVICE'S SYSTEM BOARD WAS REPLACED TO ADDRESS THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
393609 TRILOGY O2 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC. 1040004

Patients

Seq Age Sex Outcome Treatment
1