PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2015-65897
- Event Type
- Malfunction
- Date Received
- June 17, 2015
- Date of Event
- May 28, 2015
- Report Date
- May 28, 2015
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NO
- Reporter Occupation
- OTHER
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
ALL BUTTONS FUNCTIONED PROPERLY. HOWEVER, THE INSULIN PUMP HAD CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM DURING TESTING. THE INSULIN PUMP RECEIVED WITH CRACKED BATTERY TUBE THREAD, CRACKED RESERVOIR TUBE LIP, CRACKED CASE NEAR DISPLAY WINDOW CORNERS, CRACKED ON DISPLAY WINDOW, AND MINOR SCRATCHES ON DISPLAY WINDOW NOTED.
A NURSE REPORTED VIA PHONE CALL INDICATING THAT A CUSTOMER'S INSULIN PUMP ALARMED BUTTON ERROR. THE PATIENT'S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE CALL. THE NURSE WAS NOT ABLE TO ERASE THE ALARM. THE CUSTOMER SENT THE PRODUCT BACK FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 393107 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-754WWL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |