FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4850815 · Received June 17, 2015

Report

Report Number
3004209178-2015-66094
Event Type
Malfunction
Date Received
June 17, 2015
Date of Event
May 27, 2015
Report Date
May 29, 2015
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD AN INTERMITTENT BUTTON RESPONSE DUE TO FLATTENED DOWN BUTTONS DOME SWITCH. THE INSULIN PUMP HAD MINOR SCRATCHES ON LCD WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNER AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP'S DOWN ARROW WAS NOT RESPONDING PROPERLY. THE CUSTOMER STATED THAT THE DOWN BUTTON WAS STICKING. CUSTOMER DID NOT RECALL ANY SIGNIFICANT EVENTS LEADING UP TO THE ERROR ALARM. BLOOD GLUCOSE LEVEL AT THE TIME OF THE INCIDENT WAS NOT PROVIDED. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WOULD BE REPLACED AND AGREED TO RETURN THE DEVICE FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
395182 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723LNAS

Patients

Seq Age Sex Outcome Treatment
1 46 YR