PROXIMATE PROCEDURE SET
Report
- Report Number
- 3005075853-2015-03645
- Date Received
- June 17, 2015
- Date of Event
- May 15, 2015
- Report Date
- May 17, 2015
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K051301
- Removal / Correction Number
- NA
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER
Narratives
(B)(4). BREAKAWAY WASHER CUT OFF CENTER THE ANALYSIS RESULTS FOUND THAT THE PPH03 DEVICE ARRIVED IN GOOD VISUAL CONDITION; THE BREAKAWAY WASHER WAS PRESENT AND WITH AN OFF-CENTER CUT. IT APPEARS POSSIBLE THAT THE ANVIL WAS PUSHED FAR ENOUGH OFF CENTER TO RESULT IN AN OFF CENTER CUT OF THE BREAKAWAY WASHER.. THIS SITUATION NORMALLY OCCURS WHEN THE TISSUE IS NOT EVENLY DISTRIBUTED IN THE DEVICE. HOWEVER, IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE ANVIL TO BECOME OFF CENTER. IT SHOULD BE NOTED THAT ENSURING THAT THE TISSUE THICKNESS IS WITHIN THE INDICATED RANGE, AND THAT IT IS EVENLY DISTRIBUTED IN THE DEVICE. EXCESS TISSUE ON ONE SIDE MAY RESULT IN UNACCEPTABLE STAPLE FORMATION AND CAN RESULT IN STAPLE LINE LEAKAGE. PLEASE REFERENCE THE INSTRUCTIONS FOR USE FOR ADDITIONAL INFORMATION. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED AND THE DEVICE WAS TESTED FOR FUNCTIONALITY, IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND THE BATCH HAD NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING A HEMORRHOID PROCEDURE, THE STAPLER DIDN¿T FORM AND THE BLADE CUT THE TISSUE. THIS RESULTED IN BLEEDING. IT IS UNKNOWN WHAT WAS USED TO COMPLETE THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 394970 | PROXIMATE PROCEDURE SET | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |