Narratives

Additional Manufacturer Narrative · 1

(B)(4). BREAKAWAY WASHER CUT OFF CENTER THE ANALYSIS RESULTS FOUND THAT THE PPH03 DEVICE ARRIVED IN GOOD VISUAL CONDITION; THE BREAKAWAY WASHER WAS PRESENT AND WITH AN OFF-CENTER CUT. IT APPEARS POSSIBLE THAT THE ANVIL WAS PUSHED FAR ENOUGH OFF CENTER TO RESULT IN AN OFF CENTER CUT OF THE BREAKAWAY WASHER.. THIS SITUATION NORMALLY OCCURS WHEN THE TISSUE IS NOT EVENLY DISTRIBUTED IN THE DEVICE. HOWEVER, IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSED THE ANVIL TO BECOME OFF CENTER. IT SHOULD BE NOTED THAT ENSURING THAT THE TISSUE THICKNESS IS WITHIN THE INDICATED RANGE, AND THAT IT IS EVENLY DISTRIBUTED IN THE DEVICE. EXCESS TISSUE ON ONE SIDE MAY RESULT IN UNACCEPTABLE STAPLE FORMATION AND CAN RESULT IN STAPLE LINE LEAKAGE. PLEASE REFERENCE THE INSTRUCTIONS FOR USE FOR ADDITIONAL INFORMATION. THE DEVICE WAS RELOADED WITH STAPLES, A NEW WASHER WAS PLACED AND THE DEVICE WAS TESTED FOR FUNCTIONALITY, IT FIRED AND FORMED ALL THE STAPLES AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND THE BATCH HAD NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A HEMORRHOID PROCEDURE, THE STAPLER DIDN¿T FORM AND THE BLADE CUT THE TISSUE. THIS RESULTED IN BLEEDING. IT IS UNKNOWN WHAT WAS USED TO COMPLETE THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
394970 PROXIMATE PROCEDURE SET STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1