FDA Adverse Event Injury Summary report: N

CHOCOLATE PTA BALLOON

MDR report key: 4846832 · Received June 10, 2015

Report

Report Number
4846832
Event Type
Injury
Date Received
June 10, 2015
Date of Event
April 28, 2015
Report Date
June 10, 2015
Manufacturer
TRIREME MEDICAL, INC.
Product Code
LIT
Adverse Event
Yes
Product Problem
Yes
Report Source
User Facility report
Reporter Location
OH, US

Narratives

Description of Event or Problem · 1

DURING A PTA OF A RIGHT LEG A CHOCOLATE BALLOON WAS INFLATED. AFTER INFLATION THE BALLOON WAS REMOVED AND UPON INSPECTION, THE NITINOL WIRE THAT IS USUALLY ON THE BALLOON WAS NOT PRESENT. UPON FURTHER ANGIOGRAPHY OF THE RIGHT LEG THE NITINOL WIRE WAS LOCATED IN THE STENT AND WE COULD NOT RETRIEVE IT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
373509 CHOCOLATE PTA BALLOON CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT TRIREME MEDICAL, INC. * F140725007

Patients

Seq Age Sex Outcome Treatment
1 69 YR Hospitalization| O| R