FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 4844570 · Received June 11, 2015

Report

Report Number
1720753-2015-02409
Event Type
Malfunction
Date Received
June 11, 2015
Date of Event
May 20, 2015
Report Date
June 11, 2015
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE RTOS (REAL TIME OPERATING SYSTEM) CPU ASSEMBLY WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM INTERMITTENTLY FAILED TO BOOT AND EXHIBITED INTERMITTENT FUNCTIONALITY. NO PT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
381101 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900 ES-1077

Patients

Seq Age Sex Outcome Treatment
1