TERUMO ADVANCED PERFUSION SYSTEM 1
Report
- Report Number
- 1828100-2015-00491
- Event Type
- Malfunction
- Date Received
- June 11, 2015
- Date of Event
- May 20, 2015
- Report Date
- August 28, 2015
- Manufacturer
- TERUMO CARDIOVASCULAR SYSTEMS CORP.
- Product Code
- DTQ
- PMA / PMN Number
- K022947
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DURING LABORATORY EVALUATION, THE COMPLAINT WAS NOT DUPLICATED. NO ABNORMAL ALARMS OCCURED DURING TESTING OF THE AIR SENSOR AND MODULE. THE PRODUCT SURVEILLANCE TECHNICIAN (PST) PHYSICALLY AGITATED THE AIR SENSOR AND ITS CONNECTOR DURING TESTING WHICH DID NOT CAUSE FAILURE. THE PST VISUALLY EXAMINED THE AIR SENSOR WITH NOTHING OBSERVED THAT WOULD CAUSE FAILURE. THE PST TESTED THE MODULE AND SENSOR OVER A FIVE DAY PERIOD WITH NO ABNORMAL ALARMS OCCURING.
(B)(4). THE REPORTED COMPLAINT WAS CONFIRMED. DATA LOG ANALYSIS WAS PERFORMED ON THE AIR BUBBLE DETECTOR (ABD) MODULE LOGS. LOGS SHOWED MULTIPLE INSTANCES OF THE 8 VOLT (V) POWER SUPPLY TO THE ULTRASONIC AIR SENSOR FALLING OUTSIDE THE 8.0 +/- 0.4 VOLTS DIRECT CURRENT (VDC) DESIGN SPECIFICATION. LOW VOLTAGES WERE OBSERVED ON THIS LINE. DURING LABORATORY ANALYSIS, THE MODULE¿S POWER SUPPLY TO THE SENSOR WAS MEASURED TO BE 7.97 VDC WHICH IS WITHIN THE 8.0 +/- 0.4 VDC DESIGN SPECIFICATION. THE UNIT WAS RETURNED TO THE LAB FOR FURTHER TESTING. THE 8V SENSOR SUPPLY LINE WAS MONITORED WITH AN OSCILLOSCOPE SET TO TRIGGER IF THE VOLTAGE DROPPED BELOW THE 7.6 VDC DESIGN SPECIFICATION. DURING 48 HOURS OF MONITORING, THERE WERE NO ISSUES FOUND WITH THE SUPPLY VOLTAGE. SYSTEM WAS PLACED IN COLD SOAK FOR THREE HOURS THEN MONITORED. AFTER 96 HOURS OF ADDITIONAL TESTING, INCLUDING THE SYSTEM WARMING BACK UP TO AMBIENT TEMPERATURE, THE SYSTEM EXHIBITED NO VOLTAGE ANOMALIES. FROM CORRESPONDENCE WITH CUSTOMER, THE MANUFACTURING ENGINEERING CENTER (MEC) FOUND THE DEVICE WAS USED PER THE OPERATORS MANUAL, WITH THE EXCEPTION THAT IT WAS NOT TESTED BEFORE THE CASE. A LETTER WAS SENT TO MEC INSTRUCTING THEM TO REINFORCE TO THE CUSTOMER TO FOLLOW THE OPERATORS MANUAL, INCLUDING THE "TEST BEFORE USE" SECTION. NO ADDITIONAL ACTION WILL BE TAKEN AT THIS TIME.
THIS COMPLAINT IS RELATED TO MDR#: 1828100-2015-00492 AND 1828100-2015-00493. THE PERFUSION SYSTEM WAS SET-UP FOR AN URGENT DISSOCIATION SYMPTOM. THE SALES REP AND SERVICE ENGINEER CHECKED THE AIR SENSOR, BUT THE REPORTED ISSUE WAS NOT DUPLICATED. THEY CHECKED AIR SENSOR EVENT LOG. ON THE LOG, EVENT LOG "8VDC POWER SUPPLY ERROR" WAS RECORDED ON THE ISSUE'S OCCURRENCE TIME AND DATE. IT WAS THEIR FIRST TIME SEEING THIS ERROR MESSAGE.
IT WAS REPORTED THAT DURING USE OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, AFTER CARDIOPLEGIA (CPG) SOLUTION PRIMING, THE AIR BUBBLE DETECTOR (ABD) ALARMED BY ENABLING THE ABD SENSOR ON THE CENTRAL CONTROL MONITOR (CCM). THE ISSUE WAS NOT SOLVED EVEN AFTER RECONNECTING THE MODULE AND CABLE, AND THE ABD AND CABLE. THE CUSTOMER STOPPED USING THE ABD. AS A RESULT, AN ALTERNATE DEVICE WAS EMPLOYED. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 380898 | TERUMO ADVANCED PERFUSION SYSTEM 1 | APS 1 (HEART LUNG CONSOLE) | DTQ | TERUMO CARDIOVASCULAR SYSTEMS CORP. | 802110 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |