FDA Adverse Event Malfunction Summary report: N

PERFECTO2 V WITH SENSOR 9153650799

MDR report key: 4844114 · Received June 15, 2015

Report

Report Number
1031452-2015-14326
Event Type
Malfunction
Date Received
June 15, 2015
Report Date
June 18, 2015
Manufacturer
INVACARE FLORIDA OPERATIONS
Product Code
CAW
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

FOLLOW UP WILL BE FILED IF ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(6) 2015. UNIT WAS EVALUATED ON (B)(6) 2015 AND REPAIR STATEMENT SHOWS THAT THE UNIT HAD A DEFECTIVE PILOT VALVE MANIFOLD. FOLLOW UP WILL BE FILED IF ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED BY PROVIDER THAT THE RENTAL IRC5PO2V CONCENTRATOR IS SHUTTING DOWN WITHOUT ALARMING.

Description of Event or Problem · 1

(B)(6) 2015. UNIT WAS EVALUATED ON (B)(6) 2015 AND REPAIR STATEMENT SHOWS THAT THE UNIT HAD A DEFECTIVE PILOT VALVE MANIFOLD. IT WAS REPORTED BY PROVIDER THAT THE RENTAL IRC5PO2V CONCENTRATOR IS SHUTTING DOWN WITHOUT ALARMING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
387514 PERFECTO2 V WITH SENSOR 9153650799 GENERATOR, OXYGEN, PORTABLE CAW INVACARE FLORIDA OPERATIONS IRC5PO2V

Patients

Seq Age Sex Outcome Treatment
1 Other