QUICKSET ACE GRATER HEAD 54MM
Report
- Report Number
- 1818910-2015-23500
- Event Type
- Malfunction
- Date Received
- June 15, 2015
- Date of Event
- May 26, 2015
- Report Date
- May 26, 2015
- Manufacturer
- DEPUY ORTHOPAEDICS, INC.
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
EXAMINATION OF THE RETURNED INSTRUMENTS CONFIRMED THE COMPLAINT; THE GRATER HEADS ARE NOT SHARP COMPARED TO WHEN FIRST DISTRIBUTED. PROVIDED INFORMATION STATES THE INSTRUMENTS HAVE BECOME DULL OVER TIME AND NEED REPLACED. REVIEW OF THE LOT NUMBERS FOUND THE INSTRUMENTS RANGE FROM 2 TO 9 YEARS OLD. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. (B)(4).
THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS AND/OR A LOT SPECIFIC COMPLAINT DATABASE SEARCH WAS NOT POSSIBLE AS THE LOT CODES REQUIRED WERE NOT PROVIDED. PROVIDED INFORMATION STATES THE PRODUCTS HAVE BECOME DULL OVER TIME. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THE ACETABULAR REAMERS HAVE BECOME DULL OVER TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 387494 | QUICKSET ACE GRATER HEAD 54MM | HIP INSTRUMENT/TRIAL | LXH | DEPUY ORTHOPAEDICS, INC. | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |