ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2015-19720
- Event Type
- Malfunction
- Date Received
- June 14, 2015
- Report Date
- June 1, 2015
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/01/2015 WITH THE FOLLOWING FINDINGS: DEVICE EVALUATION: ON INVESTIGATION, THE DISPLAY SCREEN REMAINED BLANK WHEN THE PUMP WAS POWERED ON; THE DISPLAY SCREEN WAS CRACKED AND INOPERABLE. THE BATTERY COMPARTMENT WAS NOTED TO BE CRACKED ALONG THE SIDE AT THE THREADS AND THERE WAS MOISTURE OBSERVED INSIDE THE BATTERY COMPARTMENT.
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A DAMAGED DISPLAY SCREEN, A DAMAGED BATTERY COMPARTMENT AND A BATTERY COMPARTMENT LEAK. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 06/01/2015.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 387137 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 39 YR |