FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4842740 · Received June 12, 2015

Report

Report Number
2531779-2015-19666
Event Type
Malfunction
Date Received
June 12, 2015
Report Date
May 28, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS AT THE TIME OF THIS REPORT. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW UP #1: 09/17/2015, DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED TO ANIMAS AND EVALUATED ON 08/27/2015 WITH THE FOLLOWING FINDINGS: THE BASAL HISTORY SHOWED A 0.000U BASAL PROGRAM FROM 05/29/2015 TO 06/04/2015. THE BLACK BOX DATA FOR THESE DATES WAS NOT AVAILABLE DUE TO CONTINUED USE OF THE PUMP. THE PUMP WAS EXERCISED FOR 24HRS WITH A 1.0U/HR BASAL RATE WHICH WAS CORRECTLY RECORDED IN THE DEVICE HISTORY. THE ALLEGED COMPLAINT COULD NOT BE DUPLICATED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2015, A REPORTER CONTACTED ANIMAS ALLEGING THAT A PATIENT WAS EXPERIENCING HYPER AND HYPOGLYCEMIA. THE REPORTER DIDN'T PROVIDE ANY SPECIFIC BLOOD GLUCOSE VALUES AT THE TIME OF CONTACT, OR ANY SYMPTOMS OF BLOOD GLUCOSE EXCURSIONS. THE REPORTED ISSUE DOES NOT MEET THE CRITERIA FOR A SERIOUS INJURY. THIS COMPLAINT IS BEING REPORTED BECAUSE A PUMP MALFUNCTION COULD NOT BE RULED OUT. CUSTOMER TECHNICAL SUPPORT WAS UNABLE TO CONTACT THE REPORTER TO COMPLETE TROUBLESHOOTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
385893 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 31 YR