RUBY COIL
Report
- Report Number
- 3005168196-2015-00587
- Event Type
- Malfunction
- Date Received
- June 12, 2015
- Date of Event
- May 12, 2015
- Report Date
- May 13, 2015
- Manufacturer
- PENUMBRA, INC.
- Product Code
- HCG
- PMA / PMN Number
- K120330
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
CONCLUSION: THE DEVICE HAS BEEN RETURNED AND THE INVESTIGATION RESULTS ARE PENDING. A FOLLOW UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE DEVICE INVESTIGATION.
THE PET LOCK WAS INTACT. THE PUSHER ASSEMBLY WAS KINKED APPROXIMATELY 8.5 CM FROM THE PROXIMAL END. THE COMPLAINT HAS BEEN EVALUATED. THE COMPLAINT INDICATED THAT DURING THE PROCEDURE, THE RUBY COIL WAS DROPPED ON THE FLOOR AND WAS NOT USED FURTHER. EVALUATION OF THE RETURNED DEVICE REVEALED THE PUSHER ASSEMBLY WAS KINKED. THIS TYPE OF DAMAGE TYPICALLY OCCURS DUE TO IMPROPER HANDLING DURING USE. IF THE DEVICE IS DROPPED ON THE FLOOR DURING USE, DAMAGE SUCH AS THIS MAY OCCUR. THESE DEVICES ARE 100% VISUALLY INSPECTED AND FUNCTIONALLY TESTED DURING PROCESS INSPECTION. THE MANUFACTURING RECORDS FOR THIS LOT WERE REVIEWED AND DID NOT REVEAL ANY OUTSTANDING DISCREPANCIES, DESIGN, OR QUALITY CONCERNS.
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING RUBY COILS. DURING THE PROCEDURE, THE RUBY COIL WAS DROPPED ON THE GROUND, RENDERING IT UNSTERILE. THE PROCEDURE CONTINUED SUCCESSFULLY USING A NEW RUBY COIL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 386359 | RUBY COIL | HCG, KRD | HCG | PENUMBRA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |