SYSTEM CONSOLE
Report
- Report Number
- 1045834-2015-11323
- Event Type
- Malfunction
- Date Received
- June 12, 2015
- Date of Event
- May 13, 2015
- Report Date
- May 13, 2015
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- PK011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL NARRATIVE: THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE AND THE REPORTED CONDITION WAS CONFIRMED. AN ASSESSMENT WAS PERFORMED ON THE DEVICE WHICH FOUND THAT THE ROLLER PUMP WAS BAD DUE TO NORMAL WEAR. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE DUE TO NORMAL WEAR OVER TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
IT WAS REPORTED THAT DURING AN EAR, NOSE, AND THROAT (ENT) SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE ROLLER PUMP ON THE CONSOLE DEVICE WAS NOT SEALING PROPERLY. IT WAS REPORTED THAT THERE WAS NO DELAY DUE TO THE EVENT AS AN IDENTICAL SPARE DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE SUCCESSFULLY. THERE WAS PATIENT INVOLVEMENT REPORTED. THERE WERE NO PATIENT OR USER INJURIES REPORTED. IT WAS REPORTED THERE WAS NO MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 385414 | SYSTEM CONSOLE | MOTOR, DRILL, ELECTRIC - CONSOLE | HBC | DEPUY SYNTHES POWER TOOLS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |