FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4842209 · Received June 12, 2015

Report

Report Number
2531779-2015-19658
Event Type
Malfunction
Date Received
June 12, 2015
Report Date
May 31, 2015
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 18-JUL-2019 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THERE WERE REBOOTS OBSERVED IN THE BLACK BOX. THERE WAS NO DAMAGE FOUND TO BATTERY CAP OR BATTERY COMPARTMENT. THE BATTERY CAP WAS ABLE TO ATTACH SECURELY TO PUMP. THE PUMP POWERS ON WITH DISPLAY REMAINING BLANK. UNABLE TO PERFORM STEPS 10, 11, 13, 21 AND 35 DUE TO BLANK DISPLAY. THE BATTERY CAP CONTACTS WERE ALL FOUND TO BE WITHIN SPECIFICATION. THE PUMP WAS OPENED; NO DAMAGE FOUND TO DISPLAY SCREEN, NO DAMAGE FOUND TO POWER CIRCUIT. WHEN A TEST DISPLAY WAS USED DISPLAY FUNCTION NORMALLY. UNABLE TO ADEQUATELY INVESTIGATE REPORTED INTERMITTENT POWER DUE TO BLANK DISPLAY. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2015, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
384923 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 11 YR