FDA Adverse Event Malfunction Summary report: N

R SERIES DEFIBRILLATOR

MDR report key: 4842103 · Received June 9, 2015

Report

Report Number
1220908-2015-01442
Event Type
Malfunction
Date Received
June 9, 2015
Report Date
May 22, 2015
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K060559
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORP; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO A FAULTY TRANSFORMER OF THE PACE/DEFIB ENGINE. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED "DEFIB FAULT 76" AND "DEFIB DISABLED" MESSAGES. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
371838 R SERIES DEFIBRILLATOR DEFIBRILLATOR/ PACEMAKER MKJ ZOLL MEDICAL CORPORATION R SERIES NA

Patients

Seq Age Sex Outcome Treatment
1 NA