FDA Adverse Event
Malfunction
Summary report: N
R SERIES DEFIBRILLATOR
MDR report key: 4842103
·
Received June 9, 2015
Report
- Report Number
- 1220908-2015-01442
- Event Type
- Malfunction
- Date Received
- June 9, 2015
- Report Date
- May 22, 2015
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- MKJ
- PMA / PMN Number
- K060559
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORP; THE MALFUNCTION WAS DUPLICATED AND ATTRIBUTED TO A FAULTY TRANSFORMER OF THE PACE/DEFIB ENGINE. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT DURING FUNCTIONAL TESTING, THE DEVICE DISPLAYED "DEFIB FAULT 76" AND "DEFIB DISABLED" MESSAGES. COMPLAINANT DID NOT INDICATE THAT THERE WAS ANY PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 371838 | R SERIES DEFIBRILLATOR | DEFIBRILLATOR/ PACEMAKER | MKJ | ZOLL MEDICAL CORPORATION | R SERIES | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |